Technical details |
* Conform to CE and EN13060 |
* Manufactured under ISO 9001 and ISO 13485 norms |
* Chamber size: 247 x 460 mm, 23 Lit |
* External dimensions: 690 x 480 x 455 mm, Netweight 51 Kg |
* Power: 220 V, 50/60 Hz |
CODE | DESCRIPTION | ||||
BOE 8902300 | BOECO BTE-23D Bench Top Autoclave 23 Lit, incl 1 tray holder, 1 tray handle and 3 trays 220 V 50/60 Hz | ||||
Program | Temp | Pressure | Sterillization Time min | Total Time |
SOLID 134 | 134°C | 2.1 bar | 4 | 18 – 30 min |
SOLID 121 | 121°C | 1.1 bar | 20 | 30 – 45 min |
TEXTILE 134 | 134°C | 2.1 bar | 6 | 30 – 40 min |
TEXTILE 121 | 12°C | 1.1 bar | 20 | 35 – 50 min |
LIQUID 134 | 134°C | 2.1 bar | 10 | 45 – 65 min |
LIQUID 121 | 121°C | 1.1 bar | 30 | 50 – 75 min |
PRION | 134°C | 2.1 bar | 18 | 45 – 65 min |
B&D TEST | 134°C | 2.1 bar | 3.5 | 22 – 35 min |
HELIX TEST | 134°C | 2.1 bar | 3.5 | 22 – 35 min |
VACCUUM TEST | 15 – 20 min |
* Three times fractionated pre-vacuum |
* Stainless steel chamber |
* LCD display |
* Equipped with high quality and high-performance pump |
* 1 tray holder, 1 tray handle, 3 trays included |
* High-quality water pump and valves |
* Independent steam generator |
* Two samples self-calibration |
* Overhead type water storage tank, easy to clean |
* Drying procedure, dry by vacuum |
* Sterilization temperature: 121°C and 134°C |
* Test programs: Helix test, B&D test, and vacuum test |
* Sterilization cycles: 7 cycles are available and 3 test programs |
* Only have several buttons, very easy to use |
* All cycles are fully automatic, no need for cycle management |
* Pressure protection locking system |
* The sterilization time and dry time can be adjusted by manual |
* Malfunction detection system, an error code will be displayed |
Safety |
* Warning system prevents starting a cycle if the door is not properly locked |
* Triple threat protection for overheating and pressure |
* The pressure protection system will automatically cut off the |
electricity supply once the temperature of the chamber is abnormal. |
* Temperature control system is using a high PT1000 temperature sensor |
it is high precision, reliable service, and prompt measures. |
* The safety valve will release steam in case the pressure chamber |
exceeds normal level. |